日本日置HIOKI 3454-10数字兆欧表是日本原装进口兆欧表,却是国产兆欧表的价格。 HIOKI 3454-10数字兆欧表为50V/125V/250V/500V数字绝缘摇表 HIOKI 3454-11数字兆欧表为250V/500V/1000V数字绝缘摇表(配9294测试线) HIOKI 3454-51数字兆欧表为250V/500V/1000V数字绝缘摇表(配9289测试线) (HIOKI 9289测试线的测头都为探针,HIOKI 9294测试线的测头为一个夹子和一个探针) 
便于读数的数字显示 
250V/500MΩ 500V/500MΩ 1000V/4000MΩ 3档量程 绝缘电阻测试 
改革性的绝缘电阻测试表,具有多功能和低价格特性 
200mA持续电流 
自动放电 
活线警告 
高压测试的安全系统 
滑动机盖 
HIOKI 3454-10数字兆欧表基本参数 
| HIOKI 3454-11/HIOKI 3454-51 |  
| 绝缘测试电压  | 
250 V DC  | 
500 V DC  | 
1000 V DC  |  
| 测试量程  | 
4.000 MΩ ~ 500 MΩ | 
4.000 MΩ ~ 500 MΩ | 
4.000 MΩ ~ 4000 MΩ |  
| 第一有效量程的精度 | 
±3 %rdg. ±4 dgt. 在 0.200 ~ 50.0 M-ohm   | 
±3 %rdg. ±4 dgt. 在 0.200 ~ 50.0 M-ohm | 
±3 %rdg. ±4 dgt. 在 0.200 ~ 200.0 M-ohm |  
| hioki 3454-10 |  
| 测试电压  | 
50 V DC  | 
125 V DC | 
250 V DC | 
500 V DC |  
| 测试量程  | 
4.000 M-Ω ~ 200.0 M-Ω  | 
4.000 M-Ω ~ 200.0 M-Ω  | 
4.000 M-Ω ~ 2000 M-Ω | 
4.000 M-Ω ~ 2000 M-Ω |  
| 第一有效量程的精度 | 
±3 %rdg. ±4 dgt. at 0.200 ~ 10.00 M-ohm | 
±3 %rdg. ±4 dgt. at 0.200 ~ 10.00 M-ohm | 
±3 %rdg. ±4 dgt. at 0.200 ~ 100.0 M-ohm   | 
±3 %rdg. ±4 dgt. at 0.200 ~ 100.0 M-ohm |  
| 低电阻(连续) | 
40.00 Ω ~ 4.000 M-Ω, 短路电流:200 mA, 开路电压:4 ~ 6 V |  
| AC电压量程及精度 | 
0 ~ 750 V (50 / 60 Hz) ± 3 % rdg. ± 6 dgt., 输入阻抗 : 100 k-ohm |  
| 其他功能 | 
绝缘和低电阻模式比较器,测量值保持;绝缘电阻模式-自动放电;绝缘和AC电压-活线警告(AC70V±10V以上输入时),低电阻模式-调零;所有测量模式-电池指示,自动省电 |  
| 显示更新率 | 
2次/秒 |  
| 电源 | 
R6P锰电池×4或LR6碱性电池×4 |  
| 体积与重量 | 
175 mmW x 148 mmH x 56 mmD, 530 g  |  
| 附件 | 
9294 测试探头 测试线长约1.2m 也可购置型号﹡3454-51使用﹡9289 测试探头作为标准附件 |   
DIGITAL M-ohm HiTESTER 3454-11/3454-51/3454-10
Revolutionary insulation resistance tester with continuity function all in one low price 
- Easy-to-read digital display 
 - Auto-discharge 
 - Live-wire warning 
 - Safety system for high voltage measurement 
 - Sliding cover
  
Basic specifications 
| Model 3454-11/51 | 
250 V DC (testing voltage), 4.000 M-ohm to 500 M-ohm (measurement range), Accuracy at first effective range ±3 %rdg. ±4 dgt. at 0.200 to 50.0 M-ohm
  500 V DC (testing voltage), 4.000 M-ohm to 500 M-ohm (measurement range),  Accuracy at first effective range ±3 %rdg. ±4 dgt. at 0.200 to 50.0 M-ohm
  1000 V DC (testing voltage), 4.000 M-ohm to 4000 M-ohm (measurement range),  Accuracy at first effective range ±3 %rdg. ±4 dgt. at 0.200 to 200.0 M-ohm |  
| Model 3454-10 | 
50 V DC (testing voltage), 4.000 M-ohm to 200.0 M-ohm (measurement range),  Accuracy at first effective range ±3 %rdg. ±4 dgt. at 0.200 to 10.00 M-ohm
  125 V DC (testing voltage), 4.000 M-ohm to 200.0 M-ohm (measurement range),  Accuracy at first effective range ±3 %rdg. ±4 dgt. at 0.200 to 10.00 M-ohm
  250 V DC (testing voltage), 4.000 M-ohm to 2000 M-ohm (measurement range),  Accuracy at first effective range ±3 %rdg. ±4 dgt. at 0.200 to 100.0 M-ohm
  500 V DC (testing voltage), 4.000 M-ohm to 2000 M-ohm (measurement range),  Accuracy at first effective range ±3 %rdg. ±4 dgt. at 0.200 to 100.0 M-ohm |  
| Low resistance (continuity) | 
40.00 ohm to 4.000 M-ohm, Short circuit current:200 mA, Open terminal voltage:4 to 6 V |  
| AC voltage range | 
0 to 750 V (50 / 60 Hz) ± 3 % rdg. ± 6 dgt., Input impedance : 100 k-ohm |  
| Power supply | 
R6P(AA) manganese batteries × 4 or LR6(AA)alkaline batteries × 4 |  
| Dimensions, mass | 
175 mm(6.9 in)W × 148 mm(5.9 in)H × 56 mm(2.3 in)D, 530 g (18.7 oz) |  
| Accessories | 